Crosstex’s SporeView® Self-Contained Biological Indicators (SCBIs) for monitoring Ethylene Oxide (EO) sterilization consist of a thermoplastic vial and a green cap, a crushable glass ampule with an exclusively formulated Trypticase® Soy Broth (TSB) modified with pH indicator, and a disc inoculated with Bacillus atrophaeus spores.
- Reduced incubation time of 48 hours
- Visually distinct positive results indicated by the media transitioning from Green to Yellow
- Chemical Indicator on label provides instant indication of exposure
- Compliant with ISO 11138-1 and 11138-2
Process | EO |
Organism | Bacillus atrophaeus Cell Line 9372 |
Indications for Use | The SporView® may be utilized to monitor EO sterilization processes.This product is labeled For Industrial Use Only. |
Incubation Time | 48 hours |
Chemical Indicator | Each SCBI vial label contains a Chemical Indicator for instant indication of exposure. |
MSCBI Dimensions | 8.4 mm x 45.5 mm |
Packaging | 100 SCBIs/Box |
Compliance | ISO 11138-1 Sterilization of health care products – Biological indicators – Part 1: General requirementsISO 11138-2 Sterilization of health care products – Biological indicators – Part 2: Biological indicators for ethylene oxide sterilization processes
USP |
Shelf-Life | Typically 24 months. The shelf life of the MSCBIs is based on the shorter of the two individual components (the media ampoule and inoculated carrier), which have independent expiration periods |
Storage Conditions | 15°C to 30°C, 20-70% relative humidity |